throbber
UNITED STATES PATENT AND TRADEMARK OFFICE
`
`BEFORE THE PATENT TRIAL AND APPEAL BOARD
`
`NUVASIVE, INC.
`
`Petitioner
`
`V.
`
`WARSAW ORTHOPEDIC, INC.
`
`Patent Owner
`
`Patent Number: 8,251,997 B2
`Issue Date: August 28, 2012
`
`Case IPR2013-00206
`
`DECLARATION OF DR. JOHN W. BRANTIGAN, M.D.
`
`Mail Stop “PATENT BOARD"
`Patent Trial and Appeal Board
`US, Patent and Trademark Office
`
`PO. Box 1450
`
`Alexandria, VA 22313-1450
`
`I
`
`Nuvasive, Inc. v. Warsaw Orthopedic, Inc.
`|PR2013-00206
`
`NUVASIVE 1031
`
` 1
`
`NUVASIVE 1031
`NuVasive, Inc. v. Warsaw Orthopedic, Inc.
`IPR2013-00206
`
`

`
`I, Dr. John W. Brantigan, M.D., of Shaw Island, WA, declare that:
`
`QUALIFICATIONS
`
`1.
`
`I am an orthopaedic surgeon board certified by the American Board of
`
`Orthopaedic Surgery and fellowship trained in spine surgery.
`
`I received my medical degree
`
`from the Johns Hopkins University in 1970.
`
`I performed an internship at the Department of
`
`Surgery, University of Minnesota from 1970-1971, a residency in general surgery at the
`
`University of Utah from 1973-1974, a residency at Shriners Hospital for Crippled Children
`
`(Spokane, WA) from 1976-77, and a residency in orthopaedic surgery at the University of
`
`Washington from 19744978. From 1978 to 1990, I was in private practice as an
`
`orthopaedic surgeon at Immanuel Medical Center (Omaha, NE), Memorial Hospital of
`
`Dodge County (Fremont, NE), and St. Joseph Hospital (Omaha, NE). From 1990 to 1992, I
`
`was an orthopaedic surgeon at Cleveland Spine and Arthritis Center of Lutheran Medical
`
`Center (Cleveland, OH). From 1992 to 1997', I served as the Associate Professor of
`
`Orthopaedic Surgery and Chief of Spinal Reconstructive Surgery at Creighton University
`
`(Omaha, NE).
`
`In 1997, I returned to private practice as an orthopaedic surgeon at South
`
`Texas Orthopaedic and Spinal Surgery Associates (San Antonio, TX), after which I retired in
`
`2004.
`
`I have performed over ‘I ,000 spinal surgeries, including hundreds of surgeries
`
`involving spinal fusion implants.
`
`_ 2.
`
`I am listed as an inventor on at least? U.S. patents and additional foreign
`
`patents, including 5 U.S. patents directly related to spinal fusion implants I invented in the
`
`‘I
`
` 2
`
`

`
`1980s and early 1990s.
`
`I have published over 70 publications, including nearly 50
`
`publications pertaining to spinal implants and spinal surgical methods (many of which were
`
`published in the early 1990s).
`
`I
`
`3.
`
`I am the sole named inventor of U.S. Patent No. 5,192,327 to Brantigan (“my
`
`‘327 patent") (Ex. NUVASIVEiO06), which is addressed by the declaration of Dr. Barton L.
`
`Sachs (Ex. WARSAW2038) submitted in support of the Patent Owner‘s Responses
`
`regarding U.S. Patent No. 8,251,997 (Cases IPR2013—00206 and |PR20i3—O0208).
`
`4.
`
`Based upon my knowledge and experience in this field, I am aware of the
`
`needs and the challenges orthopaedic surgeons face in performing spinal fusion
`
`procedures.
`
`I routinely performed and observed these spinal fusion procedures, and I am
`
`familiar with the various types of spinal fusion implants that were used in 1995 and earlier,
`
`and indeed I personally invented and then used a number of spinal fusion implants during
`
`the 19805 and early 1990s.
`
`5.
`
`I am not, and never was, an employee or clinical research consultant for
`
`Nuvasive, Inc.
`
`I have been engaged in the present matter to provide my independent
`
`analysis of the issues raised by the declaration of Dr. Barton L. Sachs, and in particular
`
`statements in Dr. Sachs’ declaration about my ‘327 patent and my prior testimony.
`
`I
`
`received no compensation for this declaration beyond mynormal hourly compensation
`
`based on my time actually spent studying the matter, and I will not receive any added
`
` 3
`
`

`
`compensation based on the outcome of the abovementioned inter partes review of the ‘997
`
`patent.
`
`6.
`
`The statements in my declaration are based on my education, experience,
`
`and background in the fields discussed above and upon my review of the documents cited
`
`by Dr. Sachs in his declaration.
`
`in this declaration I do not attempt to address all of Dr.
`
`Sachs’ statements and no inference should be drawn from my not addressing a particular
`
`statement.
`
`7.
`
`In paragraph 102, Dr. Sachs begins his assessment of my ‘327 patent by stating that
`
`it “discloses ring-like prosthetic devices for anterior insertion and hemi-oval ring prosthetic
`
`devices for posterior insertion.” I disagree that the implants taught by my ‘327 patent are
`
`limited to anterior or posterior insertion. As discussed in greater detail in the following
`
`paragraphs, the implants of my ‘327 patent are, as correctly described in the patent, for
`
`“insert[ion] either anteriorly, posteriorly or laterally into the vertebral column while mounted
`
`on the end of an insertion tool." Ex. NUVAS|VE1006 at 2:64-66. See also 5:30-35 (“The
`
`individual plugs or the stack of plugs can be introduced anteriorly, laterally or posteriorly
`
`depending upon conditions and the tool receiving recesses 13 and 23 of the plugs 11 and
`
`21 can thus be positioned to meet the particular type of insertion into the vertebral
`
`column"); 6:61—69 (“This stretches the remaining disc tissue and as illustrated in FIGS. 13
`
`and 14, the plugs of this invention such as the plugs 11 or a stack of the plugs, are inserted
`
`into the opened up disc space such as 70 between adjacent vertebrae 71 and 72, either
`
`3
`
` 4
`
`

`
`anteriorly, laterally or posteriorly while mounted on a tool 73 having a single end 73a
`
`threaded into the internally threaded hole 13 of the plug 11 as illustrated in FIG. 13.")
`
`8.
`
`In paragraph 103, Dr. Sachs states, “Brantigan "327 does not disclose an implant
`
`inserted in a direct lateral path." I disagree. My ‘327 patent describes lateral insertion of an
`
`implant three times (cited above) and illustrates one example of the lateral insertion of an
`
`oval implant in Figure 10.
`
`9.
`
`In paragraph 103, Dr. Sachs states “the only reference to “lateral” insertion of an
`
`implant in Brantigan "327" is at 2:64-66.
`
`I disagree. Lateral insertion of the disclosed
`
`-implant is also described at 5:30-35 and 6:61-69 and Figure 10.
`
`10.
`
`In paragraph 103, Dr. Sachs states that I used the word "laterally" in my "327 patent
`
`to mean “something much different than ‘direct |atera|'." I disagree.
`
`In the context of the
`
`‘327 patent, “lateral” means a lateral approach to the disc space as opposed to an anterior
`
`or posterior approach as explained in greater detail below. Dr. Sachs goes on to argue in
`
`paragraph 103 that because I used the term "lateral" in other publications and in my ‘757
`
`patent to mean "to the side of the mid-sagittal line" that I used the term to mean the same
`
`thing in my ‘327 patent. That is not the case.
`
`In my ‘757 patent, I describe the implants as
`
`“inserted laterally
`
`of the vertebral column." Ex. 2042 at 1:54—55. This is different from my
`
`‘327 patent, which describes the implant as being “inserted
`
`laterally into the vertebral
`
`column" - a direction of insertion. Ex. NUVASNE1006 at col. 2:64—66; 5:30-35 (“can be
`
`introduced
`
`laterally").
`
`In the context of my ‘757 patent, the phrase “laterally of the
`
`4
`
` 5
`
`

`
`vertebral column” refers to the final resting position of the implants, not their direction of
`
`insertion, which is posteriorly.
`
`11.
`
`in paragraph 104, Dr. Sachs elaborates on his argument that the word “latera|ly" in
`
`my ‘327 patent means “to the side of the mid~sagittal line,” here contending that as used in
`
`Figure 4 of my ‘757 patent, “|aterally" means “a posterior approach that is lateral to the mid-
`
`sagittal line.” I disagree that the usage of “laterally" in my ‘757 patent applies to how the
`
`word is used in my ‘327 patent. Laterally in the context of the ‘757 patent has one meaning
`
`and in the context of the ‘327 patent, a different meaning. This difference in meaning is
`
`apparent if one considers the use of the word “lateral” in the context of the ‘327 patent.
`
`If
`
`“laterally” in the ‘327 patent meant “posteriorly," as Dr. Sachs contends, then the phrases
`
`which recite “anteriorly, posteriorly or laterally” in the ‘327 patent would be redundant and
`
`would make no sense; they would read "inserted either anteriorly, posteriorly or laterally
`
`[posterior|y].” See Ex. l\lUVAS|VE1006 at 2:64—66.
`
`12.
`
`In paragraph 104, Dr. Sachs references Fig. 31-7 in Exhibit WARSAW2055, which is
`
`a book that I edited. Fig 31-7 does illustrate a Transforaminal Lumbar lnterbody Fusion
`
`(TLIF) procedure in which two implants, which are embodiments the ‘757 patent, are
`
`inserted posteriorly. Notably, the author does not describe this approach as a lateral
`
`approach, but instead uses the phrase “bilateral” to describe the fact that the implants are
`
`placed parallel to one another via a posterior approach and lateral to the vertebral column.
`
` 6
`
`

`
`“Bi|ateral" does not describe the direction of the surgical insertion approach, which is
`
`posterior.
`
`13.
`
`In paragraph 104, Dr. Sachs references figure 12 of my ‘327 patent. He points out
`
`that “[t]he threaded insertion hoie 13 is located on the anterior side of the plug, which
`
`indicates the plug was inserted from an anterior approach.” Dr. Sachs thus contradicts his
`
`previous position that lateral means inserted using a posterior approach. Notably, my ‘327
`
`patents description of figure 12 makes no mention of the insertion direction.
`
`instead, it
`
`describes the ability of the hemioval implants to interdigitate when stacked, the use of a bar
`
`15 to hold the stack upright, and that the hollow interior of the implants are filled with bone
`
`graft material. Ex. NUVASlVE1006 at 6:41-59.
`
`14.
`
`In paragraph 104, Dr. Sachs references my testimony on cross examination during a
`
`trial between Nuvasive and Warsaw. Ex. WARSAW2041 at 1522:21—-1523:1. He states
`
`that my testimony “confirmed this interpretation of ‘laterat.”' While it is unclear which
`
`interpretation Dr. Sachs is referencing, to the extent he contends it means that I defined
`
`‘‘lateral'’ to mean a posterior or anterior insertion to the side of the rnid—saglttal plane, my
`
`testimony, which immediately follows the testimony Dr. Sachs references is directly
`
`contrary. Ex. WARSAW2041 at 1523:7510 (“Q. Okay. The mid—sagittal piane. So that's
`
`something that crosses the vertebral body like that. Now, you've defined lateral to mean to
`
`the side of that line; correct? A. No.")
`
`In fact, I testified that my use of ‘‘lateral" in the ‘757
`
` 7
`
`

`
`patent was specific to the context of my ‘757 patent, not that it was applicable to my later-
`
`filed ‘327 patent. Ex. WARSAW2041 at 1524:17—19.
`
`15.
`
`In paragraph 105, Dr. Sachs writes that I “testified that the figures [figures 10 and 11
`
`of my ‘327 patent] are not drawn correctly.’’ This earlier testimony was not referencing that
`
`Figure 10 does not show a laterally inserted implant, but instead my comment was
`
`specifically directed to the fact that the transverse processes in these illustrations “stick out
`
`more than [they] should." Ex. WARSAW2041 at 155726-12.
`
`I don‘t agree that this makes
`
`“the teachings and disclosures [of my ‘327 patent] difficult to understand," as alleged by Dr.
`
`Sachs in paragraph 105.
`
`16.
`
`In paragraph 105, Dr. Sachs states that “figures 10 and 11 of Brantigan are intended
`
`to show the same implanted plug.” Dr. Sachs then suggests that these two figures are
`“inconsistent?” I disagree. The upper portion of figure 1t) of my ‘327 patent shows a pair of
`
`my oval implants interdigitated and inserted laterally. In the lower portion of figure 10 of my
`‘327 patent we see a z—plasty annulotorny (58). This incision is used to gain access to the
`
`disc space from an anterior approach. Thus, figure 10 shows both a lateral approach
`
`(upper) and an anterior approach (lower). Figure 11 of my ‘327 patent illustrates the lower
`
`intervertebrai space of figure 10 (prepared for an anterior insertion version of my implant),
`
`not the upper portion of figure 10 that shows the laterally inserted implants. This is
`
`demonstrated by the correlation of the element numbers (52 (vertebrae), 55 (vertebral disc),
`
`56 (disc space) and 57 (lower vertebrae) between figures 10 and 11. See Ex.
`
`7
`
` 8
`
`

`
`NUVASNE1006 at 635-40.
`
`if figure 11 was showing another view of the pair of oval
`
`implants in figure 10, figure 11 would not show a single implant.
`
`lfflgure 11 was referencing
`
`the upper vertebrae, it would it identify vertebrae 51 and 52 rather than 52 and 57, it would
`
`identify disc space 54 rather than 56 and it would not identify vertebral disc 55 because in
`
`the upper space “all of the disc has been removed from the disc space 54." 6:20-21. A
`
`side—by—side comparison of FIGS. 10-11 reveals these facts:
`
`fig-9“
`
`FIG. 10
`
`I recognize that the Section Line XI-XI in figure 10 is incorrect and misplaced, but I believe
`
`the following is clear from the two figures: (a) Figure 11 is not a perpendicular cross-
`
`sectional view of what is shown in Figure 10, as is clear from the fact that Figure 11 shows
`
`the same anterior-to~posterior (front-to-back) view that is shown in the lower portion of
`
` 9
`
`

`
`Figure 10, as illustrated by the transverse processes of vertebra 57 being shown also in
`
`Figure 11 extending to the sides; and (b) Figure 11 shows the bottom two vertebrae 52 and
`
`57 of Figure 10 (with 50 being labeled in both figures as the entire spinal column, or all
`
`vertebrae in combination), and not the top two vertebrae 51 and 52 of Figure 10.
`
`17.
`
`In paragraph 106, Dr. Sachs contends that “[t]he implant or ‘plug’ depicted in Figures
`
`10 and 11 is the same plug that is depicted in Figures 1, 4, 8, 10, and 11.” I disagree. The
`
`figures show an illustrative variety of configurations of a modular design that can
`
`accommodate an anterior or posterior or lateral surgical approach. For example, figure 8 is
`
`different from figure 1.
`
`In figure 1, the bridge (or connecting bar 15, shown in Figure 3)
`
`inserts parallel to the shorter axis of the oval implant and parallel to the direction of inserter
`
`hole 13. This design is shown in a stacked configuration in figure 4.
`
`in figure 8, the
`
`connecting bar 15 is perpendicular (not parallel) to the inserter hole 13. Thus, in contrast to
`
`figure 4, the slots He in figure 8 are parallel to the longer axis of the oval and the inserter
`
`hole is on the longer axis rather than the shorter axis of the oval. A spinal surgeon having
`
`ordinary skill during the early 1990s (and even today) would have recognized that the
`
`implant illustrated in figure 8 is analogous to the implant illustrated in the upper disc space
`
`of figure 10, while the implant illustrated in figure 1 is analogous to the implant in the lower
`
`disc space of Fig 10 as shown in detail in figure 11.
`
`18.
`
`In paragraph 108, Dr. Sachs argues that “if the dashed lines [on the left side of the
`
`implants in figure 10] indicate ‘tool engagement mechanisms,’ the implant was likely
`
`9
`
`10
`
`
`10
`
`

`
`inserted anteriorly and then rotated ninety degrees per the teachings of Dr. Brantigan." I
`
`disagree. Dr. Sachs misinterprets the statement in my ‘327 patent that “the oval shape
`
`simplifies the surgical procedure since it can be rotated or reversed and still fit the
`
`vertebrae." NUVAStVE‘l006 at 2:24-25. This statement does not refer to rotation of the
`
`implant after it is inserted into the disc space.
`
`Instead, it means the implant is versatile
`
`enough to be flipped over or spun ‘I80 degrees before insertion into the patient. Because
`
`my ‘327 patent implant is symmetrical (with the exception of the inserter tool hole, which
`
`some surgeons prefer not to use), the design simplifies the surgical procedure because it
`
`makes it difficult to incorrectly insert the implant as there is no defined upper/lower surface
`
`or frontlback to the implant. Moreover, it would be anatomically impossible to rotate the
`
`implant of figure 10, even if doing so were physically possible, after placement in the patient
`
`without a significant risk of damage to the great vessels andlor spinal cord.
`
`If the implant of
`
`figure 10 were small enough to be rotated in the body as Dr. Sachs suggests, it would then
`
`be too small to be generally shaped and sized to conform with the disc space.
`
`NUVASNE1006 at 4:5—8. As discussed below, such a small size would be contrary to the
`
`disclosure of my ‘327 patent.
`
`19.
`
`In paragraph 110, Dr. Sachs argues that “blood capiilaries in a vertebral body are
`
`concentrated in the center of the vertebral body where bone tends to be more cancellous." I
`
`agree with this point, which is why my '327 patent implant has large apertures that overlay
`
`the cancellous region of the vertebral body to facilitate bone ingrowth. See, e. g., Ex.
`
`10
`
`11
`
`
`11
`
`

`
`NUVASlVE1006, col. 5:40-43. However, I disagree with Dr. Sachs’ contention that ‘“[i]f the
`
`plug 11 did rest on the edges of the implant [sic “vertebral body"), the blood capillaries
`
`would not grow into the grooves on the upper and lower surfaces of the implant." As
`
`described in column 5, lines 22-29, the ridges of my implant would “bite into” the “hard end
`
`faces” of the vertebral bodies so that the valleys would provide “gaps or troughs" for the
`
`ingrowth. During such a surgery in a live patient, there was blood residing in some or all of
`
`the valleys between the ridges of the implant, and those valleys were adjacent to the bone
`
`graft material in the central aperture of the implant, thereby leading to capillary and bone
`
`ingrowth into some or all of the “gaps or troughs” defined by the valleys. See
`
`NUVASlVE1006, col. 5:22~29.
`
`. As such, when the outer perimeter edges of the implant of
`
`my ‘327 patent rests on that cortical bone it can accommodate bone ingrowth while also
`
`reserving the more cancellous region of the vertebral body for alignment with the large
`
`fusion apertures of the implant. Of course, my ‘327 patent implant rests on this outer region
`
`of the bone because it is denser than the cancellous bone of the vertebral body and can
`
`therefore support greater mechanical loads.
`
`20.
`
`In paragraph 111, Dr. Sachs disagrees “that the length of the plugs in Brantigan ’327
`
`is sized to occupy substantially the full transverse width of the vertebral bodies and is
`
`greater than the depth of the disc space." I disagree with Dr. Sachs. While Dr. Sachs cites
`
`to my trial testimony to support this argument that my ‘327 patent does not rest on the
`
`outermost cortical ring edge of the vertebra (Ex. WARSAW2041 at 1520:9—10), he ignores
`
`11
`
`12
`
`
`12
`
`

`
`my subsequent testimony which directly contradicts his main assumption. Ex.
`
`WARSAW2041 at 1521 :1 1-21 (“Q. Okay. It says, "the periphery of the oval ring is grooved
`
`to accommodate ingrowth of blood capillaries." Do you see that? A. Yes. Q. Soto a person
`
`of ordinary skill in the art reading this, we see that figure 1, this implant, is designed to fit
`
`snuggly within the cortical ring; correct? A. No. Q. Well, you just testified that capillaries
`
`don't grow out of the apophyseal ring, did you not? A. You are mischaracterizing human
`
`anatomy")
`
`In fact, as discussed above, capillaries do grow out of the cortical bone on the
`
`perimeter of the vertebral body (because without capillaries in the cortical bone, it would be
`
`nothing more than dead matter).
`
`21.
`
`In paragraph 112, Dr. Sachs argues "that an implant that is sized and shaped to
`
`conform with an endplate would not span substantially the full transverse width of the
`
`vertebral body." I disagree. My patent teaches that “[t]he device 10 is an oval ring plug 11
`
`generally shaped and sized to conform with the disc space between adjoining vertebrae
`
`in a vertebral column." Ex. 1006 at 4:5—8 (emphasis added). Human intervertebral discs rest
`
`on the hard bone faces of the vertebrae. Given that my ‘327 patent implant was designed to
`
`replace the disc with an implant of the same size and shape, 1 disagree with Dr. Sachs’
`
`suggestion that my ‘327 patent is limited to implants substantially smaller than the disc
`
`space. My ‘327 patent never says the implant should sit on the cancellous bone as Dr.
`
`Sachs suggests; it describes just the opposite. Ex. NUVASNE1006 at 1:68-2:4 (“These
`
`ring-like prosthetic devices are bottomed on the hard bone faces or end plates of adjacent
`
`12
`
`13
`
`
`13
`
`

`
`vertebrae and are generally oval shaped to conform with the general outline perimeter of the
`
`vertebrae”).
`
`22.
`
`In paragraph 113, Dr. Sachs repeats his arguments about what “lateral" means in my
`
`‘327 patent and his argument that my ‘327 patent only discloses implants that can be
`
`inserted in anterior and posterior approaches.
`
`l disagree with both of these arguments as
`
`discussed above. ‘The term “lateral" does not mean “anterior" in my ‘327 patent, as
`
`suggested in paragraph 113 by Dr. Sachs.
`
`It means a ‘‘lateral‘' approach. The implants can
`
`be inserted in a lateral approach and my ‘327 patent specifically says so, which I believe is
`
`clear from figure 10 and reading 2:64-66, 5:30-35, 6:61-69.
`
`23.
`
`I hereby declare that all statements made herein of my own knowledge are true and
`
`that all statements made on information and belief are believed to be true; and further that
`
`these statements were made with the knowledge that willful false statements and the like so
`
`made are punishable by fine or imprisonment, or both, under Section 1001 of the Title 18 of
`
`the United States Code and that such willful false statements may jeopardize the validity of
`
`the application or any patents issued thereon.
`
`Dated: March 10, 2014
`
`W
`
`K7
`
`
`
`13
`
`14
`
`
`14

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket