throbber

`
`
`
`
`
`
`
`
`
`69’ National
`l Formulary
` rmacopeia
`
`
`
`
`
`\1 \
`
`LIES. Pha
`
`\
`
`
`
`
`
`
`
`AMGEN INC.
`
`Exhibit 1015
`
`Ex. 1015 - Page 1 of 25
`
`Ex. 1015 - Page 1 of 25
`
`AMGEN INC.
`Exhibit 1015
`
`

`

`
`
`
`
`LIJS. Phar
`
`
`
`
`
`
`
`69’ National
`'
`macopeza
`
`
`
`
`
`
`
`
`
`
`Ex. 1015 - Page 2 of 25
`
`Ex. 1015 - Page 2 of 25
`
`

`

`
`
`2000
`
`THE UNITED STATES PHARMACOPEIA
`US P 24
`
`NF 19
`THE NATIONAL FORMULARY
`
`
`
`By authority of the United States Pharmacopeial
`Convention, Inc., meeting at Washington, D. C.,
`March 9—12, 1995. Prepared by the Committee of
`Revision and published by the Board of Trustees
`
`Ofi‘ieial from January 1, 2000
`
`UNITED STATES PHARMACOPEIAL CONVENTION, INC.
`
`12601 Twinbrook Parkway, Rockville, MD 20852
`
`Ex. 1015 - Page 3 of 25
`
`
`
`r—r
`
`Ex. 1015 - Page 3 of 25
`
`

`

`
`
`
`The inclusion in the Phannaéopeia or in the National Formulary of a monograph on any drug in
`respect to which patent or trademark rights may exist shall not be deemed, and is not intended
`as, a grant of, or authority to exercise, any right or privilege protected by such patent or
`trademark. All such rights and privileges are vested in the patent or trademark owner, and no
`other person may exercise the same without express permission, authority, or license secured
`from such patent or trademark owner.
`
`NOTICE AND WARNING
`
`Concerning U.S. Patent or Trademark Rights
`
`Concerning Use of USP or NF Text
`Attention is called to the fact that USP and NF text is fully copyrighted. Authors and others
`wishing to use portions of the text should request permission to do so from the Secretary of the
`USPC Board of Trustees.
`
`The United States Pharmacopeial Convention, Inc.
`© 1999
`12601 Twinbrook Parkway, Rockville, MD 20852.
`All rights reserved
`ISSN 0195-7996
`ISBN 1-889788-03-1
`
`Printed by National Publishing, Philadelphia, PA
`
`Ex. 1015 - Page 4 of 25
`
`Ex. 1015 - Page 4 of 25
`
`

`

`USP 24
`
`Physical Tests / (788) Particulate Matter in Injections
`
`1971
`
`(788) PARTICULATE MATTER
`IN INJECTIONS
`’
`
`*
`
`Particulate matter consists of mobile, randomly—Seurced, extra—
`neous substances, other than gas bubbles, that cannot be quantitated
`by chemical analysis due to the small amount of material that it
`represents and to its heterogeneous compOsition.
`Injectable solu-
`tions, including solutions constituted from sterile solids intended
`for parenteral use, should be essentially free from particles that can
`be observed on visual inspection. The tests described herein are
`physical tests performed for the purpose of enumerating subvisible
`extraneous particles within specific size ranges.
`Microscopic and light obscuration procedures for the determi-
`nation of particulate matter are given herein.
`It is expected that
`most articles will meet the requirements on the basis of the light
`obscuration test alone; however, it may be necessary to test some
`articles by the light obscuration test followed by the microscopic
`test to reach a conclusion on conformance to requirements.
`All large-volume injections for single-dose infusion and those
`small—volume injections for which the monographs specify such re-
`quirements are subject to the particulate matter limits set forth for
`the test being applied, unless otherwise specified in the individual
`monograph.
`Not all injection formulations can be examined for particles by
`one or both of these tests. Any product that is not a pure solution
`having a clarity and a viscosity approximating those of water may
`provide erroneous data when analyzed by the light obscuration
`counting method. Such materials may be analyzed by the micro—
`scopic method. Emulsions, colloids, and liposomal preparations are
`examples. Refer to the specific monographs when a question of test
`applicability occurs. Higher limits are appropriate for certain arti-
`cles and are specified in the individual monographs.
`In some instances, the viscosity of a material to be tested may
`be sufficiently high so as to preclude its analysis by either test
`method.
`In this event, a quantitative dilution with an appropriate
`diluent may be made to decrease viscosity, as necessary, to allow
`the analysis to be performed.
`‘
`_
`In the tests described below for large—volume and small-volume
`injections, the results obtained in examining a discrete unit or group
`of units for particulate matter cannot be extrapolated with certainty
`to other units that remain untested. Thus, statistically sound sam-
`pling plans based upon known operational factors must be devel-
`oped if valid inferences are to be drawn from observed data to
`characterize the level of particulate matter in a large group of units.
`Sampling plans should be based on consideration of product vol-
`ume, numbers of particles historically found to be present in com-
`parison to limits, particle size distribution of particles present, and
`variability of particle counts between units.
`
`LIGHT OBSCURATION PARTICLE COUNT
`TEST
`
`USP Reference Standards (1 l)——USP Particle Count RS.
`The test applies to large-volume injections labeled as containing
`more than 100 mL, unless otherwise Specified in the individual
`monograph.
`It counts suspended particles that are solid or liquid.
`This test applies also to single-dose or multiple-dose small-volume
`injections labeled as containing 100 mL or less that are either in
`solution or in solution constituted from sterile solids, where a test
`for particulate matter is specified in the individual monograph. In-
`jections packaged in prefilled syringes and cartridges are exempt
`from these requirements, as are products for which the individual
`monograph specifies that the label states that the product is to be
`used with a final filter.
`
`Test Apparatus
`
`The apparatus is an electronic, liquid-borne particle counting sys—
`tem that uses a light-obscuration sensor with a suitable sample-
`feeding device. A variety of suitable devices of this type are com-
`mercially available.
`It is the responsibility of those performing the
`test to ensure that the operating parameters of the instrumentation
`are appropriate to the required accuracy and precision of the test
`
`result, and that adequate training is provided for those responsible
`for the technical performance of the test.
`It is important to note that for Pharmacopeial applicationsvthe
`ultimate goal is that the particle counter reproducibly size and count
`particles present in the injectable material under investigation. The
`instruments available range from systems where calibration and
`other components of standardization must be carried out by manual
`procedures to sophisticated systems incorporating hardware- or soft-
`ware-based functions for the standardization procedures. Thus, it
`is not possible to specify exact methods to be followed for stan-
`dardization of the instrument, and it is necessary to emphasize the
`required end result of a standardization procedure rather than a spe-
`cific method for obtaining this result. This section is intended to
`emphasize the criteria that must be met by a system rather than
`specific methods to be used in their determination.
`It is the res-
`ponsiblity of user to apply the various methods of standardization
`applicable to a specific instrument. Critical operational criteria con-
`sist of the following.
`‘

`'
`'
`Sensor Concentration Limits—Use an instrument that has a con-
`centration limit (the maximum number of particles per mL) iden-
`tified by the manufacturer that is greater than the concentration of
`particles in the test specimen to be connted. The vendor-certified
`concentration limit for a sensortis specified as that count level at
`which coincidence counts due to simultaneous presence of two or
`more particles, in the sensor view volume comprise less than 10%
`of the counts collected for lO-um particles.
`Sensor Dynamic Range—The: dynamic range of the instrument
`used (range of sizes of particlesthat can'be accurately sized and
`counted) must include the smallest particle size to be enumerated
`in the test articles.
`
`Instrument Standardization
`
`The following discussion of instrument standardization empha-
`sizes performance criteria rather than specific methods for calibrat-
`ing or standardizing a‘given instrument system. This approach is
`particularly evident in the description of calibration, where allow-
`ance must be made for manual methods as well as those based on
`firmware, software, or the use of electronic testing instruments. Ap-
`propriate instrument qualification is essential to performance of the
`test according to requirements. Since different brands of instru-
`ments may be used in the test, the user is responsible for ensuring
`that the counter used is operated according to the manufacturer’s
`specific instructions; the principles to be followed to ensure that
`instruments operate within acceptable ranges are defined below.
`The following information for instrument standardization helps
`ensure that the sample volume accuracy, sample flow rate, particle
`size response curve, sensor resolution, and count accuracy are ap-
`propriate to performance of the test. Conduct these procedures at
`intervals of not more than six months.
`'
`.
`
`SAMPLE VOLUME ACCURACY
`
`Since the particle count from a sample aliquot varies directly with
`the volume of‘ fluid sampled, it is important that the sampling ac-
`curacy is known to be within a certain range. For a sample volume
`determination, determine the dead (tare) volume in the sample
`feeder with filtered distilled or deionized water that has been passed
`through a filter having a porosity of 1.2 am or finer. Transfer a
`volume of filtered distilled or deionized water that is greater than
`the sample volume to a container, and weigh. Withdraw through
`the sample feeding device a volume that is appropriate for the spe-
`cific sampler, and again weigh the container; Determine the sample
`volume by subtracting the tare volume from the combined sample
`plus tare volumes; Verify that the value obtained is within 5% of
`the appropriate sample volume for the test. Alternatively, the sam-
`ple volume may be determined using a suitable Class A graduated
`cylinder (see Volumetric Apparatus (31)).
`[NOTE—Instruments of
`this type require a variable tare volume.’ This is the 'amount of
`sample withdrawn prior to counting. This volume may be deter-
`mined for syringe-operated samplers by setting the sample volume
`to zero and initiating sampling, so that the only volume of solution
`drawn is the tare. Subtract the tare volume from the total volume
`of solution drawn in the sampling cycle to determine the sample
`volume]
`
`
`
`
`Ex. 1015 '- Page 5 of 25
`
`
`Ex. 1015 - Page 5 of 25
`
`

`

`1972
`
`(788)
`
`Particulate Matter in Injections / Physical Tests
`
`USP 24
`
`SAMPLE FLOW RATE ‘
`
`SENSOR RESOLUTION
`
`Verify that the flow rate is within the manufacturer’s specifi-
`cations for the sensor used. This may be accomplished by using a
`calibrated stop watch to measure the time required for the instru-
`ment to withdraw and count a specific sample volume (i. e., the time
`between beginning and ending of the count cycle as denoted by
`instrdment indicator lights or other means). Sensors may be oper-
`ated accurately over a range cf flow rates. Perform the Test Pro-
`cedure at the same flow rate as that selected for calibration of the
`instrument.
`
`CALIBRATION
`
`'Use one of the following methods.
`Manual Method—Calibrate the instrument with a minimum of
`three calibrators, each consisting of near-monosize polystyrene
`spheres having diameters of about 10, 15, and 25 turn, in an aqueous
`vehicle.*, The calibrator spheres must have a mean diameter of
`within 5% of the 10-, 15—, and 25—um nominal diameters and be
`standardized against materials traceable to NIST standard reference
`materials. The number of spheres counted must be within the sen-
`sor’s concentration limit.
`.Prepare suspensions of the calibrator
`spheres in water at a concentration of 1000 to 5000 particles per
`mL, determine the channel setting that corresponds to the highest
`count setting for the sphere distribution. This is determined by
`using the highest count threshold. setting to split the distribution
`into two bins containing equal numbers of counts, with the instru-
`ment set in the differential count mode (moving window half-count
`method). Use only the central portion of the distribution in this
`calculation to avoid including asymmetrical portions of the peak.
`The portion of the distribution, which must be divided equally, is
`the count window. The window is bounded by threshold settings
`that will define a threshold voltage window of +20% around the
`mean diameter of the test spheres. The window is intended to in-
`clude all single spheres, taking into account the standard deviation
`of the spheres and the sensor resolution, while excluding noise and
`aggregates of spheres. The value of 20%‘was chosen based on the
`worst-case sensor resolution of 10% and the worst-case standard
`deviation of the spheres of 10%. Since the thresholds are propor—
`tionalito the area of the spheres rather than the diameter, the lower
`and upper voltage settings are determined by the equations:
`
`VL = 0.64V,
`
`in which V,_ is the lower voltage setting and V is the voltage at the
`peak center, and
`
`VU = 1.44v,,
`
`in which VU is the upper voltage setting.
`Once the center peak thresholds are determined, use these thresh-
`olds for the standards to create a regression of log voltage versus
`log particle size, from which the instrument settings for the 10- and
`25-um sizes can be determined.
`Automated Method—The calibration (size response) curve may
`be determined for the instrument-sensor system by the use of val-
`idated software routines offered by instrument vendors; these may
`be included as part of the instrument software or used in conjunc-
`tion with a microcomputer interfaced to the counter. The use of
`these automated methods is appropriate if the vendor supplies writ—
`ten certification that the software provides a response curve equiv-
`alent to that attained by the manual method and if the automated
`calibrationrs validated as necessary by the user.
`Electronic Method—Using a multichannel peak height analyzer,
`determine the center channel of the particle counter pulse response
`for each standard suspension. This peak voltage setting becomes
`the threshold used fOr calculation of the voltage response curve for
`the instrument. The standard suspensions to be used for the cali—
`bration are run in. order, and median pulse voltages for each are
`determined. These thresholds are then used to generate the size
`response curve manually or via software routines. The thresholds
`determined from the multichannel analyzer data are then transferred
`to the counter to complete the calibration.
`If this procedure is used
`with a comparator—based instrument, the comparators of the counter
`must be adjusted accurately beforehand.
`
`The particle size resolution of the instrumental particle counter
`is dependent upon the sensor used and may vary with individual
`sensors of the same model. Determine the resolution of the particle
`counter for IO-um particles using the monosized IO—um calibrator
`spheres. The relative standard deviation of the size distribution of
`the standard particles uSed is not more than 5%. Acceptable meth-
`ods of determining particle size resolution are ‘(1) manual deter-
`mination of the amount of peak broadening due to instrument re-
`sponse; (2) using an electronic method of measuring and sorting
`particle sensor voltage output with a multichannel analyzer; and (3)
`automated methods.
`
`Manual Method—Adjust the particle counter to operate in the
`cumulative mode or total count mode. Refer to the calibration curve
`obtained earlier, and determine the threshold voltage for the 10—p.m
`monosize spheres. Adjust 3 channels of the counter to be used in
`the calibration procedure as follows:
`Channel I is set for 90% of the threshold voltage.
`Channel 2 is set for the threshold voltage.
`Channel 3 is set for 110% of the threshold voltage.
`Draw a sample through the sensor, observing the count in Channel
`2. When the particle count in that channel has reached approxi—
`mately 1000, stop counting, and observe the counts in Channels 1
`and 3. Check‘ to see if the Channel 1 count and the Channel 3
`count are 168 i 10% and 32 i 10%, respectively, of the count
`in Channel 2.
`If not, adjust Channel 1 and Channel 3 thresholds
`to meet these criteria. When these criteria have been satisfied, draw
`a sample of suspension through the counter until
`the counts in
`Channel 2 have reached approximately 10,000, or until an appro-
`priate volume (e. g., 10 mL) of the sphere suspension has been
`counted. Verify that Channel 1 and Channel 3 counts are 168+ 3%
`and 32+ 3%, respectively, of the count in Channel 2.
`Record the particle size for the thresholds just determined for
`Channels 1, 2, and 3. Subtract the particle size for Channel 2 from
`the size for Channel 3. Subtract the particle size for Channel
`1
`from the size for Channel 2. The values so determined are the
`observed standard deviations on the positive and negative side of
`the mean count for the 10-um standard. Calculate the percentage
`of resolution of the sensor by the formula:
`
`100 [WWW
`
`in which S, is the highest observed standard deviation determined
`for the sphere, S, is the supplier’s reported standard deviation for
`the spheres, and D is the diameter, in pm, of the spheres as specified
`by the supplier. The resolution is not more than 10%.
`Automated Method—Software is available for some counters
`that allows for the automated determination of sensor resolution.
`This software may be included in the instrument or used in con-
`junction with a microcomputer interfaced to the counter. The use
`of these automated methods is appropriate if the vendor supplies
`written certification that the software provides a resolution deter-
`mination equivalent to the manual method and if the automated
`resolution determination is validated as necessary by the user.
`Electronic Method—Record the voltage output distribution of
`the particle sensor, using a multichannel analyzer while sampling a
`suspension of the 10— um particle size standard. To determine res-
`olution, move the cursor of the multichannel analyzer up anddown
`the electric potential scale from the median pulse voltage to identify
`a channel on each side of the 10- um peak that has approximately
`61% of the counts observed in the center channel. Use of the
`counter size response curve to convert the mV values of these two
`channels to particle sizes provides the particle size at within 1 stan-
`dard deviation of the 10- um standard Use these values to calculate
`the resolution as described under Manual Method.
`
`PARTICLE COUNTING ACCURACY
`
`Determine the particle counting accuracy of the instrument, using
`Method 1 (for small—volume injections) or Method 2 (for large-
`volume injections).
`
`Ex. 1015:13age 6 of 25
`
`
`Ex. 1015 - Page 6 of 25
`
`

`

`Physical Tests / (788) Particulate Matter in Injections
`
`1973
`
`USP 24
`
`Method 1—
`
`Procedure—Prepare the suspension and blank using the USP
`Particle Count RS. With the instrument set to count in the cumu-
`lative (total) mode, collect counts at settings of not less than 10 um
`and not less than 15 um. Mix the blank by inverting 25 times
`within 10 seconds, and degas the mixture by sonicating (at 80 to
`120 watts) for about 30 seconds or by allowing to stand. Remove
`the closure from the container, and gently stir the contents by hand-
`swirling or by mechanical means, taking care not to introduce air
`bubbles or contamination. Stir continuously throughout the analy—
`sis. ‘Withdraw directly from the container three consecutive vol-
`umes of not less than 5 mL each, obtain the particle counts, and
`discard the data from the first portion.
`[NOTE—Complete the pro—
`cedure within five minutes] Repeatthe procedure, using the sus-
`pension in place of the blank. From the averages of the counts
`resulting from the analysis of the two portions of the suspension at
`not less than 10 um and from the analysis of the two portions of
`the blank at not less than 10 um, calculate the number of particles
`in each mL by the formula:
`"
`.
`
`(P: ‘ Pb)/V.
`
`-
`
`in which P, is the average particle count obtained from the suspen—
`sion, P,, is the average particle count obtained from the blank, and
`V is the average volume, in mL, of the 4 portions tested. Repeat
`the calculations, using the results obtained at the setting of not less
`than 15 um.
`Interpretation—The instrument meets the requirements for Par-
`ticle Counting Accuracy if the count obtained at not less than 10
`um and the ratio of the counts obtained at not less than 10 pm to
`those obtained at not less than 15 um‘conform to the values that
`accompany the USP Particle Count RS. If the instrument does not
`meet the requirements for Particle Counting Accuracy, recalibrate
`with the remaining suspension and blank. ,If the results of the sec-
`ond test are within the limits given above, the instrument meets the
`requirements of the test for Particle Counting Accuracy.
`If on the
`second attempt the system does not meet the requirements of the
`test, determine and correct the source of the failures, and retest the
`, instrument.
`Method 2—
`
`Procedure—Using standard calibrator spheres having a nominal
`diameter of 15 to 30 um, prepare a suspension containing between
`50 and 200 particles per mL. Degas the suspension by sonicating
`(at 80 to 120 watts) for about 30 seconds or by allowing to stand.
`Properly suspend the particles by stirring gently, and perform five
`counts on 5-mL volumes of the suspension, using the particle
`counter lO-um size threshold. Obtain the mean cumulative particle
`count per mL. Pipet a volume of this suspension containing 250
`to 500 particles into a filter funnel prepared as described for Fil-
`tration Apparatus under Microscopic Particle Count. After drying
`the membrane, count the total number of standard spheres collected
`on the membrane filter. This count should be within 20% of the
`mean instrumental count per mL fer the suspension.
`
`in filtered distilled or deionized water, using a hand-held pressure
`nozzle with final filter or other appropriate filtered water source,
`such as distilled or deionized water passed through a capsule filter.
`The filter used should have a porosity of 1.2 pm or finer.
`To collect blank counts, use a cleaned vessel of the type and
`volume representative of that to be used in the test. Place a 50—mL
`volume of filtered distilled or deionized water in the vessel, and
`agitate the sample in the cleaned glassware by inversion or swirling.
`Degas by sonicating (at 80 to 120 watts) for about 30 seconds or
`by allowing to stand. Swirl the vessel containing the water sample
`by hand or agitate by mechanical means to suspend particles. »With-
`draw and obtain the particle counts for three consecutive samples
`of not less than 5 mL each, disregarding the first count.
`If more
`than 10 particles of 10 um or greater size, or more than 2 particles
`of 25 am or greater size are observed in the combined 10-mL sam—
`ple, the environment is not suitable for particulate analysis:
`the
`filtered distilled or deionized water and glassware have not been
`properly prepared or the counter is generating spurious counts.
`In
`this case, repeat the preparatory steps until conditions of analysis
`are suitable for the test.
`'
`
`Test Procedure
`
`TEST PREPARATION
`
`Prepare the test specimens in the following sequence. Outside
`of the laminar enclosure, remove outer closures, sealing bands, and
`any loose or shedding paper labels. Rinse the exteriors of the con-
`tainers with filtered distilled or deionized water as directed under
`Test Environment. Protect the containers from environmental con—
`tamination until analyzed. Withdraw the contents of the containers
`under test in a manner least likely to generate particles that could
`enter the sample. Contents of containers with removable stoppers
`may be withdrawn directly by removing the closures. Sampling
`devices having a needle to penetrate the unit closure may also be
`employed. Products packaged in flexible plastic containers maybe
`sampled by cutting the medication or'administration port tube or a
`corner from the unit with a suitably cleaned razor blade or scissors.
`Dry or lyophilized products may be constituted either by remov-
`ing the closure to add diluent or by injecting diluent with a hypo-
`dermic syringe having a 1.2-p.m'or finer syringe filter. If test spec-
`imens are to be pooled, remove the closure and empty the contents
`into a clean container.
`
`The number of test specimens must be adequate to provide a .
`statistically sound assessment of whether abatch or other large
`group of units represented by the test specimens meets or exceeds
`the limits.
`If the volume in the container is less, test a solution
`pool of 10 or more units. Single small—volume injection units may
`be tested if the individual unit volUme is 25 mL or more. For large—
`volume injections,»single units are tested. For large-volume injec-
`tions or for small-volume injections where the individual unit vol-
`ume is 25 mL or more, fewer than 10 single units may be tested,
`based on the definition of an appropriate sampling plan.
`
`Test Environment
`
`PRODUCT DETERMINATION
`
`Perform the test in an environment that does not contribute any
`significant amount of particulate matter. Specimens must be
`cleaned to the extent that any level of extraneous particles added
`has a negligible effect on the outcome of the test. The test speci-
`men, glassware, closures, and other required equipment preferably
`are prepared in an environment protected by high-efficiency partic-
`ulate air (HEPA) filters. Nonshedding garments and powder—free
`gloves preferably are worn throughout the preparation of samples.
`Cleanse glassware, closures, and other required equipment, pref-
`erably by immersing and scrubbing in warm, nonionic detergent
`solution. Rinse in flowing tap water, and then rinse again in flowing
`filtered distilled or deionized water. Organic solvents may also be
`used to facilitate cleaning.
`[NOTE—These steps describe one way
`to clean equipment; alternatively, particulate-free equipment may
`be obtained from a suitable vendor.] Finally, rinse the equipment
`
`Depending upon the dosage form being tested, proceed as di—
`rected under the appropriate category below.
`Liquid Preparations—
`Where the volume in the container is less than 25 mL—Prepare
`the containers as directed under Test Preparation. Mix and suspend
`the particulate matter in each unit by inverting the unit 20 times.
`[NOTE—Because of the small volume of some products, itmay be
`necessary to agitate the solution more vigorously to suspend the
`particles properly.]
`In a cleaned container, open and combine the
`contents of 10 or more units to obtain a volume of not less than 20
`mL. Degas the pooled solution by sonicating for about 30 seconds
`or by allowing the solution to stand undisturbed until it is free from
`air bubbles. Gently stir the contents of the container by hand-swirl—
`ing or by mechanical means, taking care not to introduce air bubbles
`or contamination. Withdraw a minimum of three aliquots, each not
`
`Ex. 1015 - Page 7 of 25
`
`
`
`
`Ex. 1015 - Page 7 of 25
`
`

`

`1974
`
`(788) Particulate Matter in Injections /‘ Physical Tests
`
`USP 24
`
`less than 5 mL in volume, into the light obscuration counter sensor.
`Discard the data from the first portion.
`'
`.Where the volume in the container is 25 mL or. more—Prepare
`the containers as directed under Test Preparation. .Mix and suspend
`the particulate matter in each unit by inverting the unit 20 times.
`Degas the solution by sonicating for about 30 seconds or by allow-
`ing the solution to stand undisturbed until it is free from air bubbles.
`Remove the closure of the unit or effect entry by other means so
`that the counter probe can be inserted into the middle of the solu-
`‘tion. Gently stir the contents of the unit by hand-swirling or by
`mechanical means. Withdraw not less than three aliquots, each not
`less than 5 mL in volume, into the light obscuration counter sensor.
`Discard the data from the first portion.
`Dry or Lyophilized Preparations—Prepare the containers as di-
`rected under Test Preparation. Open each container, taking care
`not to contaminate the opening or cover. Constitute as directed
`under Test Preparation, using the specified volume of filtered water
`or an appropriate filtered diluent if water is not suitable. Replace
`the closure, and manually agitate the container sufficiently to ensure
`dissolution of the drug.
`[NOTE—For some dry or lyophilized prod—
`ucts, it may be necessary to let the units stand for a suitable interval,
`and then agitate again to effect complete dissolution] After the
`drug in the constituted sample is completely dissolved, mix and
`suspend the particulate matter present in each unit by inverting it
`20 times prior to analysis. Proceed as directed for the appropriate
`unit volume under Liquid- Preparations, and analyze by withdraw-
`ing a minimum of three aliquots, each not less than 5 mL in volume,
`into the light obscuration counter sensor. Discard the data from the
`first portion.
`Products Packaged with Dual Compartments Constructed to
`Hold the Drug Product and a Solvent in Separate Compart-
`ments—Prepare the units to be tested as directed‘under Test Prep-
`aration. Mix each unit as directed in the labeling, activating and
`agitating so as to ensure thorough mixing of the separate compo-
`nents and drug dissolution. Degas the units to be tested by soni-
`cating or by allowing the solution to stand undisturbed until it is
`free from any air bubbles. Proceed as directed for the appropriate
`unit volume under Liquid Preparations, and analyze by withdraw-
`ing a minimum of three aliquots, each not less than 5 mL in volume,
`into the light obscuration counter sensor. Discard the data from the
`first portion.
`‘
`Products Labeled “Pharmacy Bulk Package Not for Direct
`Infusion”—Proceed as directed for Liquid Preparations where. the
`volume is 25 mL or more. Calculate the test result on a portion
`that is equivalent to the maximum dose given in the labeling. For
`example, if the total bulk package volume is 100 mL and the max-
`imum dose volume is 10 mL, then the average light obscuration
`particle count per mL would be multiplied by 10 to obtain the test
`result based on the lO-mL maximum dose.
`[NOTE—For the'cal-
`culations of test results, consider this maximum dose portion to be
`the equivalent of the contents of one full container.]
`
`Calculations
`
`Injections)—Average the
`(Small-volume
`Pooled Samples
`counts from the two or more aliquot portions analyzed. Calculate
`the number of particles in each container by theformula:
`
`PV/Van,
`
`,
`
`in which P is the average particle count obtained from the portions
`analyzed, Vr is the volume of pooled sample, in mL, Va is the vol—
`ume, in mL, of each portion analyzed, and n is the number of
`containers pooled.
`‘
`'
`’
`*
`Individual Samples (Small-volume Injections)—~Average the
`counts obtained for the 5-mL or greater aliquot portions from each
`separate unit analyzed, and calculate the number of particles in each
`container by the formula:
`'
`
`PV/V.1:
`
`in which P is the average particle count obtained from the portions
`
`analyzed, V is the volume, in mL, of the tested unit, and V, is the
`volume, in mL, of each portion analyzed.
`Individual Unit Samples (Large-volume Injections)—Average
`the counts obtained for the two or more 5-mL_ aliquot portions taken
`from the solution unit. Calculate the number of particles in each
`mL of Injection taken by the formula:
`
`' P/V,
`
`in which P is the average particle ‘countfor an individual 5 mL or
`greater sample volume, and V is the volume, in mL, of the portion
`taken.
`For all types of product, if the tested material has been diluted
`to effect a decrease in viscosity, the dilution factor must be ac-
`counted for in the calculation of the final test result.
`
`Interpretation
`
`The injection meets the requirements of the, test if the average
`number of particles present in the units tested does not exceed the
`appropriate value listed in Table 1.
`If the average number of par-
`ticles exceeds the limit, test the article by the Microscopic Particle
`Count Test.
`
`Table 1. Light Obscuration Test Particle Count.
`210 pm
`225 um
`
`Small-volume
`Injections:
`Large-volume
`Injections:
`
`6000
`
`25
`
`600 per container
`.
`
`3 per mL
`
`MICROSCOPIC PARTICLE COUNT TEST
`
`The microscopic particulate matter test may be applied to both
`large—volume and small-volume injections. This test enumerates
`subvisible, essentially solid, particulate matter in these products on
`a per-volume or per-container basis, after collection on a micro-
`porous membrane filter. Some articles cannot be tested meaning-
`fully by light obscuration.
`In such cases, individual monographs
`specify only this microscopic assay. Solutions exempted from anal-
`ysis using the microscopic assay are identified on a monograph
`basis. Examples are solutio

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket